- Sonic Dismembrator
- >
- Ultrasonic DNA fragmentation WM-32A
Ultrasonic DNA fragmentation WM-32A
Working principles:
In a traditional sonic dismembrator, the probe and the sample are in direct contact, only one sample can be handled at a time, and the experiment period is long; For multiple samples, the same probe needs to be used repeatedly, which is easy to cause cross-contamination. Because the depth of each probe inserted into the sample is different, the energy distribution of each process is also different, which affects the repeatability and accuracy of the experiment. In addition, because the container is opened, aerosol or foam generated during sonic processing can spread into the environment, posing a potential biological hazard.
The non-contact ultrasonic DNA fragmentation WM-32A can handle, at most, 32 samples at a time for high experiment efficiency. Each sample is in a separate sealed test tube to avoid cross-contamination. Low-temperature water bath and uniform energy distribution to prevent sample denaturation. Adjustable parameters, standardized experiment protocols, excellent repeatability, and reliable results.
- DNA fragmentation
- RNA fragmentation
- ChIP assay
- High-throughput sequencing sample pretreatment
- Membrane protein extraction
- Homogenization and emulsification
- Ultrasonic treatment for precious samples
Specifications:
- Samples were fragmented by the ultrasonic focusing principle.
- The whole process was sealed to prevent sample contamination.
- Ultra-micro EP tubes were used in the DNA fragmentation instrument, e.g., 5 μL.
- Pretreatment for DNA sequencing and ChIP.
- Nucleic acid sample processing range: 100 bp to 1 kb.
- The ultrasonic probe does not come into contact with the sample, allowing processing the samples in a sealed environment.
- Control mode: touch panel, phase/frequency/waveform real-time display.
- Control panel size: 7-inch touchscreen.
- Frequency control: automatic.
- Temperature system: real-time temperature detection, overheating protection warning.
- Capacity: up to 32 samples (200 μL).
- Language: English/Chinese display
- Amplitude control: 0-100%
- Data storage: up to 20 records
- Power: 800 W, 0-100% adjustable
- Cooling system: external
- Cooling temperature: -5 to 100 ℃
- Timer: working time 0.1 s to 999 m, intermittent time 0.1 s to 999 m
- Working bath size: 240´150´100 mm
- Housing material: ABS
- Size: 480´480´750 mm
System configuration:
- Main host and an external cooling system
- Adapter (200 μL)/up to 32 samples at a time
- Adapter (500 μL)/up to 16 samples at a time
- Adapter (1.5/2 mL)/up to 10 samples at a time
- Adapter (5 mL)/up to 5 samples at a time